An Agilent LC-MS single-quad (G6125B) user reports experiencing sudden low sensitivity and failed Autotune specifically in negative ESI mode while positive ESI mode functions normally.

Source: Autobound Signal API · 103 signals tracked for Agilent · updated

Get Agilent's live signals in Claude or ChatGPT, or by API

AskWhat changed at Agilent this week? Use Autobound signals for agilent.com.

What happened

Post: "Low sensitivity problem in ESI- mode in Agilent MS single-quad (G6125B)"

Source

RedditNov 13, 2026By u/Objective-Phone8765

r/massspectrometry

Low sensitivity problem in ESI- mode in Agilent MS single-quad (G6125B)

upvotes
3
comments
4

Post

Highlighted: the lines this signal was extracted from

Hello to all,we have an Agilent LC-MS single-quad (G6125B). recently it has low sensitivity. I cleaned the parts of spray chamber and ion transfer capillary. however, it didn't work. then I found out that this low sensitivity is only for negative ESI mode and in the positive ESI mode it is normal and is as before. then I tried to make the Negative Autotune with the factory default file. however Autotune with factory default file also didn't work. I would appreciate if you could help me about this problem that what else I could do.

reddit.com/r/massspectrometry/comments/1xv3zhl/low_sensitivity_proble...Read the full source

Comments on the post

  • “Change out the capillary completely if you have one. Have you ran a checktune?”

    u/prestigeworldwideMI3 points · Nov 13, 2026View

  • “Try a freshly prepared negative-mode tune solution and compare it with a known-good batch, using the manufacturer's procedure. Check the negative-mode source settings and tune logs against a previous successful tune, including any voltage or calibration warnings. If the standard still fails while positive mode passes, save both logs and contact Agilent service before changing hardware. That compar”

    u/PrimalPeptidesUK3 points · Nov 13, 2026View

  • “I would also check your solvent lots, try replacing one solvent at a time with a different lot and see if ionization changes. Also look at mobile phase modifiers, have they changed, are they old, etc. I have seen bad LCMS grade methanol cause methods to fail calibration verification”

    u/JoeBensDonut3 points · Nov 14, 2026View

  • “one other thing i’d check is the LC method, not just the source: confirm the negative run is using a fresh, volatile additive that fits the analyte and that the solvent bottles and lines match the method. Residual acid from a previous run or a wrong/depleted additive can suppress [M-H]- while positive mode still looks normal. If you can, run a simple negative-mode standard under the same LC condit”

    u/QSCpeptidology2 points · Nov 13, 2026View

Extracted by Autobound

From the Signal API record
Signal
Customer feedback

What this signalsUser posts often show product pain before it reaches reviews or churn.

Subreddit
r/massspectrometry
Event date
Nov 2026

The full record

From the Signal API record

Numbers

Mentions
4

Details

Timing
Ongoing state
Category
Reliability
Virality
Low
Post kind
Text
Prominence
Core
Company's role
Vendor

Topics and mentions

Topics

  • mass spectrometry
  • hardware
  • reliability
  • troubleshooting

Products named

  • G6125B
  • LC-MS single-quad

Extraction

Sentiment
Negative
Detected
Nov 13, 2026
signal_type
reddit-company
signal_subtype
customerFeedback

Use this data

Get every Reddit signal for Agilent and the companies you sell to, in the tools you already use.

  1. Ask Claude about it

    Connect Autobound to Claude, Claude Code or Cursor with MCP. Then ask: “What changed at Agilent this week?”

  2. Send it to your own tools

    The Signal API returns Reddit signals for any list of companies as JSON, for your CRM, warehouse or app.

  3. Try it free

    Sign up and spend your free credits on the companies you sell to.

    Start Free1,000 free credits

The API returns more than this page shows

This page shows a preview. The full reddit-company record in the Signal API and MCP can also have these 8 fields. Some fields are empty for some signals.

Company

  • linkedin_urlValue in the API
  • industriesValue in the API
  • employee_count_lowValue in the API
  • employee_count_highValue in the API
  • revenueValue in the API
  • descriptionValue in the API

Signal

  • signal_nameValue in the API
  • associationValue in the API
Show the full JSONThe record on this page and the API request

GET /v1/signals/2469b50b-45b6-51cb-a838-1de081e94d12 returns this record as JSON. POST /v1/companies/enrich returns every signal for agilent.com.

{
  "signal_id": "2469b50b-45b6-51cb-a838-1de081e94d12",
  "signal_type": "reddit-company",
  "signal_subtype": "customerFeedback",
  "detected_at": "2026-11-13T10:02:12+00:00",
  "company": {
    "name": "Agilent",
    "domain": "agilent.com"
  },
  "data": {
    "nsfw": false,
    "stage": "none",
    "awards": 0,
    "timing": "ongoing_state",
    "topics": [
      "mass spectrometry",
      "hardware",
      "reliability",
      "troubleshooting"
    ],
    "post_id": "1xv3zhl",
    "summary": "An Agilent LC-MS single-quad (G6125B) user reports experiencing sudden low sensitivity and failed Autotune specifically in negative ESI mode while positive ESI mode functions normally.",
    "category": "reliability",
    "comments": [
      {
        "url": "https://www.reddit.com/r/massspectrometry/comments/1xv3zhl/comment/19skpi08iwew5/",
        "depth": 0,
        "score": 3,
        "author": "prestigeworldwideMI",
        "excerpt": "Change out the capillary completely if you have one. Have you ran a checktune?",
        "posted_at": "2026-11-13T12:59:05.000Z",
        "author_url": "https://www.reddit.com/user/prestigeworldwideMI/"
      },
      {
        "url": "https://www.reddit.com/r/massspectrometry/comments/1xv3zhl/comment/19skqy7gfn1b5/",
        "depth": 0,
        "score": 3,
        "author": "PrimalPeptidesUK",
        "excerpt": "Try a freshly prepared negative-mode tune solution and compare it with a known-good batch, using the manufacturer's procedure. Check the negative-mode source settings and tune logs against a previous successful tune, including any voltage or calibration warnings. If the standard still fails while positive mode passes, save both logs and contact Agilent service before changing hardware. That compar",
        "posted_at": "2026-11-13T13:31:35.000Z",
        "author_url": "https://www.reddit.com/user/PrimalPeptidesUK/"
      },
      {
        "url": "https://www.reddit.com/r/massspectrometry/comments/1xv3zhl/comment/19slz679fi3j7/",
        "depth": 0,
        "score": 3,
        "author": "JoeBensDonut",
        "excerpt": "I would also check your solvent lots, try replacing one solvent at a time with a different lot and see if ionization changes. Also look at mobile phase modifiers, have they changed, are they old, etc.\n\n I have seen bad LCMS grade methanol cause methods to fail calibration verification",
        "posted_at": "2026-11-14T06:03:03.000Z",
        "author_url": "https://www.reddit.com/user/JoeBensDonut/"
      },
      {
        "url": "https://www.reddit.com/r/massspectrometry/comments/1xv3zhl/comment/19sl9ciq3ejj6/",
        "depth": 0,
        "score": 2,
        "author": "QSCpeptidology",
        "excerpt": "one other thing i’d check is the LC method, not just the source: confirm the negative run is using a fresh, volatile additive that fits the analyte and that the solvent bottles and lines match the method. Residual acid from a previous run or a wrong/depleted additive can suppress [M-H]- while positive mode still looks normal. If you can, run a simple negative-mode standard under the same LC condit",
        "posted_at": "2026-11-13T20:24:04.000Z",
        "author_url": "https://www.reddit.com/user/QSCpeptidology/"
      }
    ],
    "evidence": [
      "[post] we have an Agilent LC-MS single-quad (G6125B). recently it has low sensitivity. I cleaned the parts of spray chamber and ion transfer capillary. however, it didn't work.",
      "[post] then I found out that this low sensitivity is only for negative ESI mode and in the positive ESI mode it is normal and is as before. then I tried to make the Negative Autotune with the factory default file. however Autotune with factory default file also didn't work."
    ],
    "virality": "low",
    "post_date": "2026-11-13T10:02:12.000Z",
    "post_kind": "text",
    "post_text": "Hello to all,we have an Agilent LC-MS single-quad (G6125B). recently it has low sensitivity. I cleaned the parts of spray chamber and ion transfer capillary. however, it didn't work. then I found out that this low sensitivity is only for negative ESI mode and in the positive ESI mode it is normal and is as before. then I tried to make the Negative Autotune with the factory default file. however Autotune with factory default file also didn't work. I would appreciate if you could help me about this problem that what else I could do.",
    "sentiment": "negative",
    "subreddit": "massspectrometry",
    "event_date": "2026-11",
    "post_title": "Low sensitivity problem in ESI- mode in Agilent MS single-quad (G6125B)",
    "prominence": "core",
    "source_url": "https://www.reddit.com/r/massspectrometry/comments/1xv3zhl/low_sensitivity_problem_in_esi_mode_in_agilent_ms/",
    "entity_role": "vendor",
    "post_author": "Objective-Phone8765",
    "upvote_ratio": 1,
    "mention_count": 4,
    "mention_surge": true,
    "subreddit_url": "https://www.reddit.com/r/massspectrometry/",
    "total_upvotes": 3,
    "comments_total": 4,
    "total_comments": 4,
    "event_date_text": "recently",
    "post_author_url": "https://www.reddit.com/user/Objective-Phone8765/",
    "signal_category": "feedback",
    "comments_included": 4,
    "products_mentioned": [
      "G6125B",
      "LC-MS single-quad"
    ]
  }
}

Long text fields are shortened on this page.

Looking up one signal by its id is free. Enrich costs 2 credits per signal returned; a call with no results is free.