A clinical laboratory reports that a new lot of Roche lipase reagent on the cobas c503 fails high QC with Bio-Rad MultiQual, displays a sharp mean shift from 144 to 90, and fails linearity testing.

Source: Autobound Signal API · 87 signals tracked for Roche · updated

Get Roche's live signals in Claude or ChatGPT, or by API

AskWhat changed at Roche this week? Use Autobound signals for roche.com.

What happened

Post: "Roche new lot of lipase issue??? What are we doing"

Source

RedditAug 27, 2026By u/Icy-Fly-4228

r/medlabprofessionals

Roche new lot of lipase issue??? What are we doing

upvotes
5
comments
18

Post

Highlighted: the lines this signal was extracted from

So I know we aren’t the only one, just wondering how you all are handling it. New lot of lipase reagent (c503) will not pass high QC( bio rad multi qual). Roche says it’s due to matrix effect. Tech support gal said other available lot is acting the same way. Roche QC works fine, Old reagent lot mean 144. New reagent lot mean 90. Bio rad has a bulletin acknowledging it’s an issue. I did further testing and it also is extremely inaccurate between reagent lots with CAP survey material and the new fails linearity with Maine standards products.

reddit.com/r/medlabprofessionals/comments/1w073fi/roche_new_lot_of_li...Read the full source

Comments on the post

5 of 18 comments
  • “Do patient comparisons between the old and new lot of reagent. If the values are within your acceptable TAE (probably 1/4 TAE), then it’s truly a matrix effect and you can adopt a new QC mean after. If they are exceeding your TAE, then it’s not a matrix effect. It’s bad reagent/calibration. But it should just be matrix effect. It happens, especially with certain tests.”

    u/Brofydog14 points · Aug 27, 2026View

  • “What lot number is it? We have a new lot we are going into soon.”

    u/glitterfae14 points · Aug 27, 2026View

  • “This is an example why you always use third party QC.”

    u/Western_Produce_86451 points · Aug 29, 2026View

  • “Yeah, we used the Preci controls after the BioRad controls were constantly out. We adjusted the ranges after that.”

    u/mauieclectic1 points · Aug 31, 2026View

  • “I will check my lot and come back to you but for me lipase is usually the strangest one. It starts to trend low like crazy and happens for any lot from my experience. For our c503 (doesn't matter which line), the cfas cal never works the first couple times until the 3rd or 4th time which passes and the qc jumps straight back into the mean for both levels. For my lab we establish our means over the”

    u/kristpy1 points · Sep 1, 2026View

Extracted by Autobound

From the Signal API record
Signal
Customer feedback

What this signalsUser posts often show product pain before it reaches reviews or churn.

Subreddit
r/medlabprofessionals

Companies

  • Bio-RadAlso named
  • Maine StandardsAlso named

The full record

From the Signal API record

Numbers

Mentions
1

Details

Timing
Ongoing state
Category
Reliability
Virality
Medium
Post kind
Text
Prominence
Core
Company's role
Vendor

Topics and mentions

Topics

  • diagnostics
  • quality control
  • laboratory
  • reagents

Flair

  • Technical

Products named

  • cobas c503
  • lipase reagent

Extraction

Sentiment
Negative
Detected
Aug 27, 2026
signal_type
reddit-company
signal_subtype
customerFeedback

Use this data

Get every Reddit signal for Roche and the companies you sell to, in the tools you already use.

  1. Ask Claude about it

    Connect Autobound to Claude, Claude Code or Cursor with MCP. Then ask: “What changed at Roche this week?”

  2. Send it to your own tools

    The Signal API returns Reddit signals for any list of companies as JSON, for your CRM, warehouse or app.

  3. Try it free

    Sign up and spend your free credits on the companies you sell to.

    Start Free1,000 free credits

The API returns more than this page shows

This page shows a preview. The full reddit-company record in the Signal API and MCP can also have these 8 fields. Some fields are empty for some signals.

Company

  • linkedin_urlValue in the API
  • industriesValue in the API
  • employee_count_lowValue in the API
  • employee_count_highValue in the API
  • revenueValue in the API
  • descriptionValue in the API

Signal

  • signal_nameValue in the API
  • associationValue in the API
Show the full JSONThe record on this page and the API request

GET /v1/signals/d6e3ccd4-10b2-507c-aaa3-6f33a7abc243 returns this record as JSON. POST /v1/companies/enrich returns every signal for roche.com.

{
  "signal_id": "d6e3ccd4-10b2-507c-aaa3-6f33a7abc243",
  "signal_type": "reddit-company",
  "signal_subtype": "customerFeedback",
  "detected_at": "2026-08-27T21:28:08+00:00",
  "company": {
    "name": "Roche",
    "domain": "roche.com"
  },
  "data": {
    "nsfw": false,
    "stage": "none",
    "awards": 0,
    "timing": "ongoing_state",
    "topics": [
      "diagnostics",
      "quality control",
      "laboratory",
      "reagents"
    ],
    "post_id": "1w073fi",
    "summary": "A clinical laboratory reports that a new lot of Roche lipase reagent on the cobas c503 fails high QC with Bio-Rad MultiQual, displays a sharp mean shift from 144 to 90, and fails linearity testing.",
    "category": "reliability",
    "comments": [
      {
        "url": "https://www.reddit.com/r/medlabprofessionals/comments/1w073fi/comment/p6b06p7/",
        "depth": 0,
        "score": 14,
        "author": "Brofydog",
        "excerpt": "Do patient comparisons between the old and new lot of reagent. If the values are within your acceptable TAE (probably 1/4 TAE), then it’s truly a matrix effect and you can adopt a new QC mean after.\n\n If they are exceeding your TAE, then it’s not a matrix effect. It’s bad reagent/calibration.\n\n But it should just be matrix effect. It happens, especially with certain tests.",
        "posted_at": "2026-08-27T22:20:26.000Z",
        "author_url": "https://www.reddit.com/user/Brofydog/"
      },
      {
        "url": "https://www.reddit.com/r/medlabprofessionals/comments/1w073fi/comment/p6bepj2/",
        "depth": 0,
        "score": 4,
        "author": "glitterfae1",
        "excerpt": "What lot number is it? We have a new lot we are going into soon.",
        "posted_at": "2026-08-27T23:35:17.000Z",
        "author_url": "https://www.reddit.com/user/glitterfae1/"
      },
      {
        "url": "https://www.reddit.com/r/medlabprofessionals/comments/1w073fi/comment/p6mldvw/",
        "depth": 0,
        "score": 1,
        "author": "Western_Produce_8645",
        "excerpt": "This is an example why you always use third party QC.",
        "posted_at": "2026-08-29T15:44:52.000Z",
        "author_url": "https://www.reddit.com/user/Western_Produce_8645/"
      },
      {
        "url": "https://www.reddit.com/r/medlabprofessionals/comments/1w073fi/comment/p6wylnx/",
        "depth": 0,
        "score": 1,
        "author": "mauieclectic",
        "excerpt": "Yeah, we used the Preci controls after the BioRad controls were constantly out. We adjusted the ranges after that.",
        "posted_at": "2026-08-31T03:22:15.000Z",
        "author_url": "https://www.reddit.com/user/mauieclectic/"
      },
      {
        "url": "https://www.reddit.com/r/medlabprofessionals/comments/1w073fi/comment/p75xqdd/",
        "depth": 0,
        "score": 1,
        "author": "kristpy",
        "excerpt": "I will check my lot and come back to you but for me lipase is usually the strangest one. It starts to trend low like crazy and happens for any lot from my experience. For our c503 (doesn't matter which line), the cfas cal never works the first couple times until the 3rd or 4th time which passes and the qc jumps straight back into the mean for both levels. For my lab we establish our means over the",
        "posted_at": "2026-09-01T12:16:22.000Z",
        "author_url": "https://www.reddit.com/user/kristpy/"
      }
    ],
    "evidence": [
      "[post] New lot of lipase reagent (c503) will not pass high QC( bio rad multi qual). Roche says it’s due to matrix effect.",
      "[post] Old reagent lot mean 144. New reagent lot mean 90. Bio rad has a bulletin acknowledging it’s an issue.",
      "[post] I did further testing and it also is extremely inaccurate between reagent lots with CAP survey material and the new fails linearity with Maine standards products."
    ],
    "virality": "medium",
    "post_date": "2026-08-27T21:28:08.000Z",
    "post_kind": "text",
    "post_text": "So I know we aren’t the only one, just wondering how you all are handling it. New lot of lipase reagent (c503) will not pass high QC( bio rad multi qual). Roche says it’s due to matrix effect. Tech support gal said other available lot is acting the same way. Roche QC works fine, Old reagent lot mean 144. New reagent lot mean 90. Bio rad has a bulletin acknowledging it’s an issue. I did further testing and it also is extremely inaccurate between reagent lots with CAP survey material and the new fails linearity with Maine standards products.",
    "sentiment": "negative",
    "subreddit": "medlabprofessionals",
    "post_flair": [
      "Technical"
    ],
    "post_title": "Roche new lot of lipase issue??? What are we doing",
    "prominence": "core",
    "source_url": "https://www.reddit.com/r/medlabprofessionals/comments/1w073fi/roche_new_lot_of_lipase_issue_what_are_we_doing/",
    "entity_role": "vendor",
    "post_author": "Icy-Fly-4228",
    "upvote_ratio": 0.78,
    "mention_count": 1,
    "mention_surge": false,
    "subreddit_url": "https://www.reddit.com/r/medlabprofessionals/",
    "total_upvotes": 5,
    "comments_total": 27,
    "total_comments": 18,
    "other_companies": [
      {
        "name": "Bio-Rad",
        "role": "partner"
      },
      {
        "name": "Maine Standards",
        "role": "partner"
      }
    ],
    "post_author_url": "https://www.reddit.com/user/Icy-Fly-4228/",
    "signal_category": "feedback",
    "comments_included": 5,
    "products_mentioned": [
      "cobas c503",
      "lipase reagent"
    ]
  }
}

Long text fields are shortened on this page.

Looking up one signal by its id is free. Enrich costs 2 credits per signal returned; a call with no results is free.