Troubleshooting Acquity FL Sample Manager
- upvotes
- 1
- comments
- 7
Post
Highlighted: the lines this signal was extracted from
My M class fixed loop sample manager is not delivering sample, every run looks like the solvent only. Detector is working (lockmass is A-OK) and I can see the increasing AcN gradient. As far as I can discern the mechanical movements are all OK and I have checked and reset the Z-height. The sample manager passes all the built in tests for seals, needle loop volume calibration, etc. By removing and using a syringe pump I have determined that the needle and loop are both clear. I have tested the inlet valve for clear passage between all 3 pairs of ports in both load and inject positions. The sample pressure during injection reads about 160 psi. The only "anomaly" is that I had to increase the air bubble size in order to get it to measure the needle and loop volumes. Any suggestions would be appreciated!